INSREF Ultra
Cryostat Circulator, Cat. No. IRI 017
Polished Stainless Steel tank of filling volume 10-12
ltrs with overall size 300 mm x 200 mm X 200 mm (D)
Large usable workspace of approximately 180 mm X 180 mm x 200 mm (D)
Pump cum stirrer with pumping capacity of between 10 & 12 ltrs /
min at zero head
Long - life steel immersion heater with extra long cold zone
All other wetted parts are of non-corrosive, non-ferrous materials
Tank insulation is foamed - in - place Polyurethane Foam (PUF)
Operating temperature range as per table below
Digital proportional, fully solid state temperature control
Accuracy of temperature control better than ± 0.1o
C at minimum temperature
Digital temperature display with resolution (least cont) of 0.1o
C
Accuracy of temperature display ± 0.5% of full rang ±1 digit
High temperature safety cut-off for heater (over temperature
protection)
Auto boost heating (if max. temperature > 40oC)
Auto above ambient cut-off for compressor (if max. temperature >
40oC)
Cooling by air cooled, balance, non CFC, single or double stage
systems
For double stage systems, necessary safety cut-offs are built-in to
allow unattended operation
Built-in high & low voltage cut-off to protect compressors from
power supply extremities
Powder coated, compact, double color exterior, with an insulated lid
Easy to operate & maintain, with detachable controller for
ease-of-recalibration
Available in two formats - floor standing (shown alongside) with
wheels for easy mobility and table top
Cat.
No.
|
Temperature
Rang,oC
|
Cooling
Capacity, Watts |
Maximum
Outer
Dimension, mm |
Table
Top |
Floor |
IRI
017 A |
-10
to 40 |
260
@ 0oC |
750
X 650
450 (H)
|
600
X 475
825 (H)
|
IRI
017 C |
-20
to 40 |
480
@ 0oC |
IRI
017 E |
-30
to 40 |
670
@ 0oC |
IRI
017 G |
-40
to 40 |
250
@ -40oC |
NOT
AVAILABLE
|
650
X 600
900 (H)
|
IRI
017 I |
-70
to 40 |
320
@ -40oC |
Pump
motor extends to an height 200 mm above top cover of the equipment.
Pump nozzles are of ID 6 mm, shoe -cut and can take flexible tubing
having up to 10 mm ID. Nozzles are approximately 30 mm above top
cover of the equipment.
Lowest temperatures typically reach within 90 to 120 minutes, at an
ambient to 27oC to 30oC. The lowest reachable
temperature varies with ambient temperature, quality & nature,
of tank media and level of equipment maintenance. An air -
conditioned environment is most suited, though not absolutely
necessary.
While calculating space requirements based on the above data,
pleased take into account free air circulation space of at least 300
mm on all sides of the equipment.
Suggested
tank liquids are distilled water for operating temperatures above 5oC
and good quality methanol for operating temperature below 5oC.
The Digital Temperature Display is calibrated at our factory against
thermometers which have calibration certificates traceable to
national standards. Calibration Certificates from a Government of
India lab, traceable to national standards for individual equipment
can be given, at extra cost.
OPTION
AVAILABLE
Smaller / larger tank sizes
Smaller frame sizes
Higher Maximum Temperatures
Higher Cooling Capacities
Stands / holdres
Calibration Certificate
Insulation Tubing
WHILE ORDERING, PLEASE INDICATE CHOICE OF TABLE TOP
OF FLOOR FORMAT.
IF THERE IS NO SUCH INDICATION, ONLY FLOOR FORMAT WILL BE SUPPLIED
|